Table 3.1 SiteStand Station Troubleshooting
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION
The system battery doesn’t
charge in the SiteStand
station.
If the charge function of the
system is acceptable, the
possible causes are:
the power supply has
failed or become
disconnected
the dock PCBA located
in the sleeve assembly
has failed
the docking connectors
on the sleeve assembly
are broken or bent.
Verify the SiteStand station
is at fault by attempting to
charge the battery with the
power supply connected
directly to the system. If it
does not charge, refer to the
SonoSite Ultrasound System
Service Manual for
troubleshooting.
There is no video from the
SiteStand station video out
connection.
The SiteStand station
docking PCBA has failed.
The docking connections or
the sleeve assembly have
been broken.
See Replacing the Sleeve
Assembly and the Docking
PCBA on page 11.
There is no AC power from
the SiteStand station power
connection.
The SiteStand station fuse
has failed.
See Replacing the SiteStand
station Fuse on page 13.
There is no remote printer
control to the printer on the
SiteStand station.
The printer remote control
cable has failed or is not
attached correctly.
Replace the printer remote
cable (P00762).
There is no output from the
RS-232 communication
connector.
The SiteStand station
docking PCBA has failed.
The docking connections on
the sleeve assembly are
broken or bent.
See Replacing the Sleeve
Assembly and the Docking
PCBA on page 11.
The sleeve assembly does
not raise or lower.
The foot pedal is damaged.
The gas spring is loose or
damaged.
The foot pedal has fallen off.
See Replacing the Foot
Pedal on page 11.
The SiteStand station does
not roll easily or in a straight
line.
A castor is broken or bent.
The SiteStand station base is
damaged.
Replace the failed part.