Datasheet

G
THERMASGARD
®
HFTM
Rev. 2017 - V12 GB
Calibratable temperature measuring transducer with sleeve sensor
THERMASGARD
®
HFTM
with eight switchable measuring ranges, continuous
output, with or without optional display (displaying the actual temperature), terminal box enclosure made of impact-resistant plastic and enclosure
cover with quick-locking screws. The temperature transmitter with remote sensor is used for temperature detection in liquid and gaseous media,
e.g. by installation inside an immersion sleeve, or as duct sensor. The sensors are factory-calibrated. Adjustment ⁄fine adjustment by the user is
possible (zero point offset is adjustable). Direct and permanent use in liquids is possible in combination with THE immersion sleeves.
TECHNICAL DATA
Power supply: 24 V AC ⁄ DC (± 10 %) for U variant
15...36 V DC for I variant, depending on working resistance, residual ripple stabilised ± 0.3 V
Working resistance: R
a
(ohm) = (U
b
-14 V) 0.02 A for I variant
Load resistance: R
L
> 5 kOhm for U variant
Power consumption: < 1.0 VA 24 V DC; < 2.2 VA 24 V AC
Sensor: Pt1000, DIN EN 60751, class B
Measuring ranges:
multi-range switching with 8 switchable measuring ranges,
see table (other ranges optional)
with manual zero point correction 10 K)
Deviation, temperature: ± 0.2 K at +25 °C
Output: 0 - 10 V or 4... 20 mA
Ambient temperature: measuring transducer –30...+70 °C
Connection type: 2- or 3-wire connection
Protective tube: stainless steel, 1.4571, V4A, Ø = 6 mm,
(sensor sleeve) nominal length NL = 50 mm (other dimensions optional)
cable entry stamped (optional rolled)
Connection cable: silicone, SiHF, 2 x 0.25 mm², KL = 1.5 m (other lengths and range limits optional,
e.g. PTFE leads up to +250 °C or
glass fibre with steel wire mesh up to +350°C)
Enclosure: plastic, material polyamide, 30 % glass-globe-reinforced,
with quick-locking screws (slotted ⁄ Phillips head combination),
colour traffic white (similar to RAL 9016), enclosure cover
for display is transparent!
Enclosure dimensions: 72 x 64 x 37.8 mm (Tyr 1 without display)
72 x 64 x 43.3 mm (Tyr 1 with display)
Cable gland: M 16 x 1.5, including strain relief, exchangeable,
max. inner diameter 10.4 mm
Electrical connection: 0.14 - 1.5 mm² via terminal screws on circuit board
Humidity: < 95 % r. H., non-precipitating air
Protection class: III (according to EN 60 730)
Protection type:
IP 65
(according to EN 60 529) humidity-tight
stamped
IP 68
(optional watertight compound-filled)
rolled
IP 54
(optional
glass fibre
cable)
Standards: CE conformity, electromagnetic compatibility according to
EN 61326, according to EMC directive 2014 ⁄ 30 ⁄ EU
Optional: Two-line
display with illumination,
cutout approx.
36 x15 mm (W x H), for displaying the
ACTUAL
temperature
and the
internal diagnostics
(measuring
range exceeded, measuring range not reached, sensor
breakage, sensor short circuit)
THERMASGARD
®
HFTM
Sleeve sensors with temperature measuring transducer
Type ⁄ WG01 Sensor Output Type
Display Item No.
HFTM - I IP 65, I - variant
HFTM-I Pt1000 4...20 mA Remote sensor 1101-1152-0219-920
HFTM-I DISPLAY Pt1000 4...20 mA Remote sensor 1101-1152-2219-920
HFTM - U IP 65, U - variant
HFTM-U Pt1000 0 - 10 V Remote sensor 1101-1151-0219-920
HFTM-U_DISPLAY Pt1000 0 - 10 V Remote sensor 1101-1151-2219-920
Extra charge: Other ranges optional, Protection type
IP 68
(Sensor sleeve watertight compound-filled)
2-wire connecting leads, per running meter (
silicone
/
PTFE
/
glass fibre
) on request
Other lengths of protection sleeve (NL) optional on request
Accessories
THE - xx Immersion sleeve
, stainless steel or nickel-plated brass, Ø = 9 mm
Display and internal diagnostics
Measuring range
exceeded
Standard
Measuring range
not reached
Sensor
breakage
Sensor
short circuit