Specifications
16
General Discussion of Calibraon
In order to have condence in the measurements taken
with a radiaon detecon instrument, a calibraon must
be performed. The calibraon compares the response
of the instrument to known values and the instrument
is adjusted (if needed) to bring all indicaons within
the range of the instrument to within ±10% of the true
values. Generally, radiaon detecon instruments should
be calibrated annually.
CAUTION: Errors can occur when using low level sources
or background for calibrang. In the Calibraon mode,
the smallest increment which can be adjusted is .010.
Pulse Calibraon
Since GM-based instruments are essenally pulse
counters, it must be veried that the number of counts
reported equals the number of counts input. This should
be demonstrated for both count rates and total counts.
The issue is confused somewhat with the addion of
dead-me compensaon where the number of counts
in any given second is adjusted upward to account
for pulses “lost” during the me the detector is busy
recharging from a pulse.
Requirements
You will need a PC running Observer USB and a Pulse generator
Procedure
1. With the instrument powered o, remove the back of the unit and aach a pulse generator to
the unlabeled test point immediately to the le of U2. Pulses must be posive with a duraon of
approximately +3.3 to 5V peak to peak square wave. No high voltage is connected to this test point.
2. Power on the instrument.
3. Connect the instrument to the PC running Observer USB using the USB cable. An applicaon window for
the aached instrument should open within a few seconds.
4. Go to View > Cal Panel menu of The Observer USB Soware to access the Calibraon Screen.
5. Record the currently displayed dead me. (Note the current deadme)
6. Change the dead me to 0 and click the Update Sengs buon. Aer a second or two the instrument
should beep to acknowledge that the new seng has been received, stored, and applied. Seng the
dead me to 0 seconds causes the device to not apply any dead-me correcon at all, so the display
should report the true input pulse rate.
7. Run the pulse generator at several dierent rates to span the range of the instrument. For the Inspector
EXP, the maximum working input pulse rate should be about 367,000 CPM (6.123 Hz). Verify that the
value reported by the instrument is within ±10% of the input value in all cases.
8. Change the dead me back to the seng recorded in Step 5.
Chapter 8: Observer USB Calibraon Soware
All calibraon factors must be entered using the
Observer USB soware.










