Specifications
10
Ulity Menu
The Ulity Menu allows the user to change default sengs for several operang parameters. Once a seng is
changed, it remains in eect unless they are changed through the Ulity Menu or the Observer USB Soware.
1. To acvate the Ulity Menu, hold down the + buon on the end panel while turning the instrument on.
The display will show the MENU icon. Release the + buon and donE will appear along with the MENU
icon. The ulity menu can be accessed even if the probe is disconnected.
2. Scroll through the menu by pushing the + or - buons.
3. To select an opon, push the SET buon and the SET icon is displayed.
4. Use the + or - buons to toggle between choices and press the SET buon to enter the new value. The
instrument will connue to operate in the Ulity Menu mode, and the display will show DONE and
MENU. To adjust another Ulity Menu opon, repeat the above steps.
5. To exit the Ulity Menu at any me, press the SET buon again when DONE is displayed. The Inspector
EXP will connue with the normal start-up roune.
Opons
Opons Funcon Comments
dLoG Data Logging (The logging
interval can be set in the USB
Observer Soware)
Turns on or o the data logging feature for
storing collected readings in the internal
memory. Default is 10 minutes.
UNIT Unit of Measurement Selects between mR/hr and CPM or μSv/hr
and CPS
EFF Eciency Displays the eciency for the pre-pro-
grammed isotopes built into the Inspector.
CLOC Clock (We recommend the clock be
set prior to collecng data to ensure
a correct me and date in your
collected data. )
Sets the internal clock for use with the data
logging feature.
dEF Default Sengs (Does not
change calibraon.)
Set the unit to the default sengs (Alarm
o, Auto Averaging on, mR/hr &CPM,
recording o, clears and resets isotopes and
backlight me) and exits the Ulity Menu
VEr Version Displays the rmware version
USE Use specic isotope eciency If surveying a known isotope, select from
the list of isotopes eciencies to display
acvity in Bq or DPM
donE Done Press the SET buon when “donE” is dis-
played to exit the menu and begin taking
readings.
AVEr Averaging Turns Auto-Averaging on and o