6 Maintenance
6.1 Periodic calibration check of the HygroClip 2 probe
Both the Pt 100 RTD temperature sensor used in the probe and associated electr on ic s are v ery stable and
should not require any calibration after the initial factory adjustment. Long term stability of the ROTRONIC
Hygromer humidity sensor is typically better than 1 %RH per year. For maximum accuracy, calibration of the
probe should be verified every 6 to 12 months. Applications where the probe is exposed to significant pollution
may require more frequ ent verific atio ns.
6.2 Cleaning or repla cing the probe dust filter
See document E-M-HC2 Probes-V1
6.3 Firmware updates
Firmware updates will be available on the ROTRONIC website for downloading. Firmware files are given a
name that shows both to which device the file applies and the version number of the firmware. All firmware
files have the extension HEX.
WARNING: for reasons of compatibility update the current instrument Firmware only with a Firmware file with
the same “Main” version. Examples: 1.0a 1.5b = OK, 1.3b 2.1 d = NOT OK
Procedure for updating the firmware:
o Connect the HygroLab C1 to a USB port of a PC with the ROTRONIC HW4 software installed. Note
that the ROTRONIC USB driver must be installed on the PC as explained in the HW4 manual E-M-
HW4v3-Main
Note: the Ethernet port can be also used to connect the HygroLab C1 to the HW4 PC.
.
o Copy the firmware update file from the ROTRONIC website to the PC.
o Start HW4 software on the PC and search for the H y groLab C1 (HW4 Main Menu Bar > Devices and
Groups > Search for USB Masters).
o After finding the HygroLab C1, expand the device tree to its functions. Select Device Manager. In the
Device Manager menu bar select Tools > Firmware Update.
For instructions see document E-M-HW4v3-F2-019