during operation, the sensor probes are checked against failure by checking the valid 
dynamic measurement ranges. All EEPROM updates, initiated by the sensor itself, as 
well as by external connections, are checked by subsequent memory read back and 
data comparisons. These different system checks return error bytes to the system 
RAM. The error codes are available by connecting a PC with a special RS232 cable 
connected to the UART port slide connector. The error codes are shown  in the 
software UIP (version 4.3 or later). Warm up and Out of Range are the only bits that 
are reset automatically after return to normal state. All other error bits have to be 
reset manually after return to normal by power off/on. 
The yellow LED flashes if an error has been detected. If a fatal error has been 
detected the yellow LED is lit. 