4. Refit the dial, place the sealing gasket on the front face of the case then centre the window on the gasket. Refit the
bezel ring tightly with firm hand pressure.
CAUTION! Observe correct safety procedures when working with high temperature calibration baths.
Figure 3 - Internal view of case showing calibration trimmers and °C / °F selection jumpers
Minimum and maximum reading option
The Digitemp may be fitted with an option which stores the maximum and minimum measured temperatures. These temperatures
may be displayed or cleared by means of magnetically-operated switches without opening the sealed case.
To display the minimum and maximum temperatures : Place a bar magnet on the top right-hand side of the case within the
box marked Min - Max on the instrument label. The DigiTemp display will then alternate between the maximum and minimum val-
ues as long as the magnet is held in place. The maximum temperature is shown after the display message Hi and the minimum
value follows Lo.
To reset (clear) the minimum and maximum values : Place a bar magnet on the top left-hand side of the case within the box
marked Min-Max Clear on the instrument label. After approximately 5 seconds the minimum and maximum values will be
cleared, confirmed by the display message HiLo CLrd. Note that the minimum and maximum readings will also be cleared if the
battery is removed.
Warranty
The Rototherm DigiTemp thermometer is guaranteed to be free of manufacturing defects for a period of one year from the date of
purchase. To obtain service under this warranty the instrument should be returned in its original packaging either to British
Rototherm or its distributors. This warranty excludes the battery, damage due to battery leakage or misuse. Any details given in
this instruction sheet are subject to change without notice.
DigiTemp Digital Thermometer
Instruction Manual
Page 3
R