User Guide

RECEIPT OF RECORDER
The Clearscan recorder complete with a packet of charts, door keys and pen pack(s) is dispatched in a protective casing
which should preferably be replaced after inspection, as protection until the instrument is ready for installation.
The spare pen pack(s) and door keys are inside the protective case.
To assist with inspection, a label attached on the inside face of the case door states instrument serial number, chart number,
range of each pen and the rotation speed of the chart drive.
Additional parts for pipe mounting versions are packed in the protective casing during transit.
RECEIPT OF TEMPERATURE RECORDER
The capillary tubing between the bulb and recorder is coiled for despatch purpose. Immediately prior to installation, this
tubing must be uncoiled carefully in order to avoid twisting or kinking.
INSTALLATION AND MOUNTING - wall or panel mounting
Ideally the site chosen should be free from dust, corrosive fumes, vibration and extremes of temperature.
The instrument is suitably designed for outdoor weather conditions and protected against ingress of dust and water.
The recorder is suitable for wall or panel mounting using the brackets provided.
Refer to figure 1 for overall dimensions of the recorder and general installation.
3
CASE BREATHER
53
56
371
41 114
490
302
420483
PRESSURE CONNECTION
1/2" NPT STANDARD
OR 1/4" NPT OPTIONAL
327
312
WALL MOUNTING
PANEL MOUNTING
PANEL CUTOUT SIZE: 305 x 425
Fig. 1a Outline Dimensions for 12 inch chart recorder (wall or panel mounting)
all dimensions are in mm