10.3. BATTERY PACK REPLACEMENT
It is recommended that the battery packs contained in each unit should be replaced
every 300 charge / discharge cycles. This number can be read and displayed with the
optional RFD Test Box. If a Test Box is not available, this number can be estimated
by your typical usage. It is recommended that the battery pack is replaced by the
factory so that the case seals can be fully tested prior to re-deployment. For this
reason, it may be desirable to arrange battery pack replacement with Scheduled
Service of the RFD system. To ensure quality and safety, only OEM battery packs
can be used in the RFD units.
Battery Pack Replacement Procedure
10.3.1. Battery pack placement may be performed by the user under certain
conditions. Battery packs replacement should only be performed indoors in a dry,
clean, dust-free work environment. Replacement should only be performed by
persons having sufficient skills and tools to perform the work described below. The
use of non OEM battery packs is not allowed.
1. Loosen and remove the 9 or 13 smaller machine screws on the backside of the
unit using a 7/64 Allen wrench. Then loosen and remove the four larger machine
screws using a 9/64 Allen wrench.
2. Separate the lid from the main body exposing the battery.
3. A black plastic connector housing at the bottom of the case attaches to the 3-wire
battery connector plug. Carefully cut the white nylon cable tie from the clip on the
battery pack and remove.