User Manual

Side 6
© ROPOX 2016
Product Information
Installation interfaces
Construction
The wall must be made of a material suited for screw mounting and approved for damp
rooms.
The lateral adjustment system has been load-tested and is safe up to 400 kg/881lbs.
According to DS/EN 12182, the maximum user weight amounts to 200 kg/441lbs.
Max. load of upper screws: 41kg/screw ( 90lbs/screw).
Electrics
Recommended location of socket in wall for optimum and hidden wiring.
In case of an alternative solution, the cable from the control unit is 3.2m. See Fig. 3.
Plumbing
The drain has a diameter of Ø32 (11/4”)and is provided with an Ø32x88.5° elbow.
Max. protrusion of drain connection and elbow from the wall is 50mm/1 19/20 owing to the
covers.
The water supply ends in 2 (two) 90° shut-off valves with ½” outside thread. Location and
optimum orientation as illustrated. Max. protrusion of shut-off valves from the wall is 50mm
owing to the covers.
It is recommended to install water and drain connections as illustrated. All hoses for water
and drain connections must be flexible for the unit to move freely within the range of
adjustment.