Specifications
24 ST-133/ST-133A Controller Manual Version 2.H
Aux. Trig. Out: Ac coupled variable delay trigger output for synchronizing
other system components with PTG. The host software sets the Delay
Time of the auxiliary trigger output with respect to the PTG trigger time.
This output does not need to be connected to PI-MAX.
Trig. Indicator: LED trigger indicator. 100 ms flash is produced each time the
PTG triggers. With repetition rates faster than 10 Hz, indicator glows
continuously.
Power Input Module: This module contains the line-cord socket, the Power On/Off
switch and two fuses. The power and fuse requirements are printed on the panel
above the module. For more detailed information, see Power Requirements on
page 12.
Software Installation
It is necessary to install the application software before the controller can be operated
and data acquired. The installation procedure will vary according to the computer type,
operating controller, and type of application software. See your software manual for
detailed software installation and software operation information.
Imaging Field of View
When used for two-dimensional imaging applications, Princeton Instruments LN/CCD,
TE/CCD and RTE/CCD cameras closely imitate a standard 35 mm camera. Since the
CCD is not the same size as the film plane of a 35 mm camera, the field of view at a
given distance is somewhat different. The imaging field of view is indicated in Figure 7.
D
O
B
S
Object
Lens
CCD
Figure 7. Imaging field of view
D = distance between the object and the CCD
B = 46.5 mm for F mount; 17.5 mm for C mount
F = focal length of lens
S = CCD horizontal or vertical dimension
O = horizontal or vertical field of view covered at a distance D
M = magnification
The field of view is:
where
O =
S
M
,
()
M
=
FD
D
–
B
2










