User manual

Table Of Contents
USB 2.0 Interface Addendum
Roper Scientific 13 of 14 June 17, 2003
E:\Manual s\USB 2. 0 Install ation\USB 2 .0 Addendum to Manual.doc
No CCD Named in the Hardware Wizard:CCD dialog
Figure 12. Hardware Wizard: Detector/Camera/CCD dialog box
If you have installed a USB 2.0 Interface Module in your ST-133A, a blank field may be
displayed in the Detector/Camera/CCD dialog box (Figure 12) if the ST-133 controller
was made before January 2001. Earlier versions of the ST-133 did not contain non-
volatile RAM (NVRAM), which is programmed with information about the controller
and the camera. PVCAM, the program under which the Roper Scientific USB works,
retrieves the information stored in NVRAM so it can enter specific camera
characteristics into WinView/32 or WinSpec/32.
Check the serial label on underside of your controller. If the first five characters are
D1200 (December 2000) or earlier (J0797 or July 1997, for example), contact Customer
Support to find out about an NVRAM controller upgrade.