Quick Start Reference – Calibrating Density
Step 1 Start at the Status Display
From the Status Display Screen, Press the F3 Key to display the
Password Screen (or the Main Menus if the password is disabled).
C
F1 F2 F3 F4
Ronan Engineering
X96S Density Gage
Enter Password:
******
All0
RST0
EXIT
ENTER
C
F1 F2 F3 F4
Ronan Engineering
X96S Density Gage
Status Display
D1 12.1860 % sol-wt
D2 1.1860 SGU
Proc Temp 50.00 deg
Raw Counts 12000
Filt.Counts 12000
Time 04/20/05 08:05
Diagnostics
BACK
Step 2 Password Menu
With the left most character * highlighted (All digits are set at
000000 at this point), enter in the password ‘101010’.
Press the ↑ key 1 time to make the left most character equal to 1.
Press the → key 2 times to move to the third character.
Press the ↑ key 1 time to make the third character equal to 1.
Press the → key 2 times to move to the fifth character.
Press the ↑ key 1 time to make the fifth character equal to 1.
Press the F4 key (enter) to accept the password.
This will take you to the Main Menu.
Note: If the wrong password was entered, press the F1 (All0) to set
all the characters back to a value of 000000. You can begin to re-
enter the password from the beginning. Pressing the F2 (RST0) will
set the individual character that is highlighted back to the value 0.
Note: For security reasons, each character will always be displayed
as an asterisk
*.