Owner manual
30
Manual Weight Input BariSelect
Wheninnormaloperatingmode,presstheManualbuttononthekeypad(fig.
30).Thecontrolunitwillentermanualweightinputmode.Thedisplaywill
change from the system pressure level to user weight in pounds, starting at 400
pounds each time the manual mode is entered.
Use the Firmer and Softer buttons to adjust the displayed weight to the
approximateweightoftheuser(fig.31).TheFirmerbuttonwillincreasethe
displaybytwenty(20)poundswitheachpress,theSofterbuttonwilldecrease
thedisplayby(10)poundswitheachpress.
Oncethedesiredweightisdisplayedandnobuttonsarepressedforfive(5)
seconds, the control unit will revert to normal operation. The display will revert
back to the system pressure level and the control unit will then adjust the
pressure to the new setpoint. Allow time for the system pressure to stabilize
before making further adjustments.
Pulsate Mode
The BariSelect control unit is equipped with the Pulsate
Mode. When selected, the control unit will continuously
readjustthesystempressure(proportionaltotheComfort
Levelsetting).Thechangeisgradualtoensuremaximum
patient comfort.
• Unlock the control unit.
• PressthePulsatekey(Fig.32).
• Select the Pulsate mode setting by pressing the up and down keys.
• Press the Pulsate key to set the mode into memory. Ensure the proper
mode is indicated on the LCD screen.
You may change the rate of pulsation at any time while the Pulsate Mode is in
operation.
•
•
•
Section 12 BariSelect
Figure 30
Manual
1
Figure 32
Figure 31