Owner manual
22
Note: To assist in deflating the mattress, disconnect the hose at the
mattress or control unit.
Thenpressthe"UprightOn/Off"buttonasecondtime(Ensuretheupright
lightextinguishes).
Note: Do not leave the control unit in the Upright Mode when the
patient is lying flat or at a slight angle.
Note: Ensurethepatienthasatleastone(1)inch(3cm)ofclearance
between the bed frame and their body. If not, manually increase or
decrease system pressure to compensate.
Firm Mattress Mode
To aid in the placement and maneuvering of the patient, the control unit is
equipped with a Firm Mattress Mode. The Firm Mattress Mode will fully inflate
the mattress making it firm for easy positioning. The Firm Mattress Mode will
automatically terminate when the time displayed on the keypad has expired.
Press the "Firm" button on the keypad. The display will indicate the
remainingtime(fig.15).
To end the cycle press the Firm Mattress “Off” button.
Note: The SelectAir control unit is equipped with an internal timer which will
turn the Firm Mattress Mode off after 20 minutes. If the Firm Mattress
Mode has timed out and ended the firm mattress cycle, DO NOT press
the Firm "Off” key as this may cause the system to "lock up".
Note: Pressing the Firm "Off" key multiple times or holding down the button
may cause the system to “lock up”.
Note: To ensure patient comfort, the control unit slowly lowers the pressure of
the system. This may give the appearance that the firm function is still
active. If the timer lights are extinguished, the control unit is in its
normal mode.
1.
2.
Section 10 SelectAir
Figure 16
SelectAir,
SelectAir Max,
HybridSelect
Models
Flashing
“Green” Light
(Powerto
ControlUnit)
CPR/Emergency Deflate
Important: In case of a CPR Emergency, turn power OFF by pressing the
CPR button on the keypad.
Figure 15