R&S®Cable Rider ZPH Cable and Antenna Analyzer Specifications Data Sheet | Version 04.
Version 04.00, December 2018 CONTENTS Definitions ....................................................................................................................................................................... 3 Specifications .................................................................................................................................................................. 4 Frequency .....................................................................................................
Version 04.00, December 2018 Definitions General Product data applies under the following conditions: Three hours storage at ambient temperature followed by 30 minutes warm-up operation Specified environmental conditions met Recommended calibration interval adhered to All internal automatic adjustments performed, if applicable Specifications with limits Represent warranted product performance by means of a range of values for the specified parameter.
Version 04.
Version 04.00, December 2018 Distance-to-fault analysis Result formats return loss, SWR, split screen DTF and SWR, split screen DTF and return loss Return loss Range 1/2/3/5/10/20/30/50/100/120/130/150 dB, linear 100 % 0.01 dB Resolution SWR Range Fault resolution Maximum cable length selectable depending on cable loss Immunity to interference Maximum permissible spurious signal 1 to 1.1/1.5/2/3/6/11/21/71 (1.5 × 108 × velocity factor/span) m 1500 m (nom.
Version 04.00, December 2018 Spectrum Analyzer (R&S®ZPH-K1 option, only model .
Version 04.00, December 2018 Level Display range Maximum rated input level DC voltage CW RF power Peak RF power Max. pulse voltage Max. pulse energy Intermodulation Third-order intercept (TOI) Second-harmonic intercept (SHI) Displayed average noise level displayed noise floor to +30 dBm duration < 3 s pulse width 10 µs intermodulation-free dynamic range, signal level 2 × –20 dBm, RF attenuation = 0 dB, RF preamplifier = off f = 1 GHz +7 dBm (meas.) f = 2.4 GHz +10 dBm (meas.
Version 04.00, December 2018 Level measurement uncertainty Absolute level uncertainty at 100 MHz Frequency response (+20 °C to +30 °C) Attenuator uncertainty Uncertainty of reference level setting Display nonlinearity Bandwidth switching uncertainty Total measurement uncertainty +20 °C to +30 °C 5 kHz ≤ f < 10 MHz 10 MHz ≤ f ≤ 4 GHz < 0.3 dB < 1.5 dB (nom.) < 1 dB < 0.3 dB < 0.1 dB (nom.) < 0.3 dB SNR > 16 dB, 0 dB to –50 dB, logarithmic level display reference: RBW = 10 kHz < 0.1 dB (nom.
Version 04.00, December 2018 R&S®ZPH-K7 analog modulation analysis AM/FM (model .12 only) Measurement of analog modulation signals Center frequency Demodulation bandwidth Bandwidth accuracy Display 10 MHz to 4 GHz 2 MHz, 1 MHz, 500 kHz, 300 kHz, 200 kHz, 100 kHz, 50 kHz, 30 kHz, 20 kHz, 10 kHz (nom.) < +/- 5% (nom.
Version 04.00, December 2018 Channel power meter (R&S®ZPH-K19 option, model .02 only) Frequency range with R&S®ZPH-B4 option installed Measurement range Measurement accuracy +20 °C to +30 °C, 2 MHz ≤ f < 10 MHz +20 °C to +30 °C, 10 MHz ≤ f ≤ 4 GHz –20 °C to +50 °C, 10 MHz ≤ f ≤ 4 GHz 2 MHz to 3 GHz 2 MHz to 4 GHz –20 dBm to +30 dBm < 2 dB (nom.) < 0.8 dB < 1.2 dB Channel power meter (R&S®ZPH-K19 option, model .
Version 04.00, December 2018 General data Manual operation Languages Chinese, Chinese Traditional, English, French, German, Italian, Hungarian, Japanese, Korean, Portuguese, Russian, Spanish Remote control Command set LAN interface USB Display Resolution Audio Speaker USB interface SCPI 1997.0 10/100BASE-T, RJ-45 mini B plug, version 2.
Version 04.00, December 2018 EMC Dimensions W×H×D Weight Recommended calibration interval in line with European EMC Directive 2014/30/EU including EN 61326-1 class B (emission) CISPR 11/EN 55011/group 1 class B (emission) EN 61326-1 table 2 (immunity, industrial) 202 mm × 294 mm × 76 mm (8.0 in × 11.6 in × 3 in) 2.5 kg (5.5 lb) 1 year Equivalence of specifications for different R&S®ZPH part numbers The specifications for part number 1321.1211.02 are equivalent to part number 1321.1211.52 and 1321.
Version 04.00, December 2018 Ordering information Designation Type Order No. Handheld cable and antenna analyzer, 2 MHz to 3 GHz R&S®Cable Rider ZPH 1321.1211.02 Handheld cable and antenna analyzer, 2 MHz to 3 GHz R&S®Cable Rider ZPH 1321.1211.
Version 04.
Version 04.00, December 2018 Optical power sensors and accessories Designation OEM USB optical power meter (Germanium) OEM USB optical power meter (filtered InGaAs) SC adapter for optical power meter LC Adapter for optical power meter 2.5 mm universal adapter for optical power meter 1.
Worldwide Local and personalized ❙ Customized and flexible ❙ Uncompromising quality ❙ Long-term dependability ❙ ❙ Rohde & Schwarz The Rohde & Schwarz electronics group offers innovative solutions in the following business fields: test and measurement, broadcast and media, secure communications, cybersecurity, monitoring and network testing.