Datasheet
Version 04.00, December 2018 
4  Rohde & Schwarz R&S
®
Cable Rider ZPH Cable and Antenna Analyzer 
Specifications  
Frequency 
Frequency range 
R&S
®
Cable Rider ZPH 
2 MHz to 3 GHz 
with R&S
®
ZPH-B4 option installed 
2 MHz to 4 GHz 
Frequency resolution 
1 Hz 
Reference frequency, internal 
Total reference accuracy 
±(time since last adjustment × aging rate) 
+ temperature drift + calibration accuracy 
Aging per year 
±1 × 10
–6
Temperature drift 
0 °C to +30 °C 
±1 × 10
–6
+30 °C to +50 °C 
±3 × 10
–6
Achievable initial calibration accuracy 
±5 × 10
–7
Measurements 
Individual 
reflection (S
11
) 
1-port cable loss 
distance-to-fault 
model .12 
transmission (S
21
) 
Measurement wizard 
Guides the user through a sequence of individual measurements. Uses the R&S
®
Instrument View PC software to configure the 
measurement sequence including hints displayed on the screen. R&S
®
Instrument View is also used to combine the measurement 
results into user-configurable reports. 
Measurement setup 
Port output power 
model .02 
–10 dBm (nom.) 
model .12 
–3 dBm to –30 dBm step 1 dB (nom.) 
Data points 
selectable 
101 to 2501 
Measurement bandwidth 
reflection measurement (S
11
) 
10 kHz 
transmission measurement (S
21
) 
10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz, 
1 MHz 
Trace modes 
clear/write, average 
Reflection measurement S
11
Result formats 
magnitude, SWR, magnitude and 
distance-to-fault, SWR and distance-to-
fault, smith chart, phase 
Magnitude 
Range 
1/2/3/5/10/20/30/50/100/120/130/150 dB, 
linear 100 % 
Resolution 
0.1 dB 
SWR 
Range 
selectable 
1 to 1.1/1.5/2/3/6/11/21/71 
Measurement speed 
0.3 ms per point 
Corrected directivity with R&S
®
ZN-Z103 
2 MHz ≤ f ≤ 4 GHz 
> 42 dB (nom.) 
Corrected test port match 
with R&S
®
ZN-Z103 
2 MHz ≤ f ≤ 4 GHz 
> 36 dB (nom.) 
Measurement uncertainty 
with R&S
®
ZN-Z103 
see figure Uncertainty of reflection 
measurement 
1-port cable loss measurement 
Result format 
magnitude 
Range 
selectable 
1/2/5/10/20/50/100 dB 
Resolution 
0.1 dB 