Chapter 3: Operation
Command FM0
command function
FM0 <FM-deviation)> FM-deviation query
The command FM0 reads the actual FM-deviation value in the VOR mode.
Example:
command from the controller
response of the VOR/ILS Analyzer EVS200
query:
FM0 <CR>
fm0 <CR>
e.g. 479.1Hz <CR/LF>
e.g. 479.1Hz <CR/LF>
Command FM1
command function
FM1 <FM-Index)> query of FM-index value
The command FM1 reads the actual FM-index value in the VOR mode.
Example:
command from the controller
response of the VOR/ILS Analyzer EVS200
query:
FM1 <CR>
fm1 <CR>
e.g. 15.9 <CR/LF>
e.g. 15.9 <CR/LF>
Command D1
command function
D1 <Display-information> switching over to Display1
The command D1 enables the switching over to the content of display1 in the VOR / ILS mode.
This is the prediction for reading out the data of the corresponding mode (VOR/ILS).
Example:
command from the controller
response of the VOR/ILS Analyzer EVS200
input:
D1 <CR>
d1 <CR>
READY <CR/LF>
READY <CR/LF>