User manual

VOR/ILS Analyzer EVS200
0796.1800.02 E-8 33
Chapter 3: Operation
Command RES
command function
RES <RESET> Master-RESET
With the command RES a unit master reset is possible. It works like a restart of the unit.
Example:
command from the controller response of the VOR/ILS Analyzer EVS200
input:
RES<CR>
res<CR>
READY <CR/LF>
READY <CR/LF>
EVS200 REMOTE-SYSTEM READY <CR/LF>
Command BI
command function
BI <BITE-test information> BITE (Build In Test) -asking for result
The EVS200 performs approx. every 500 ms a self test (BITE) and makes the result available as a
9bit binary form. During this test the following voltages and funtionally subunits of the units are
checked.
all operating voltages,
hostprocessor
memory,
DSP-processor,
battery capacity (%)
explanation of BITE information
B8 B7 B6 B5 B4 B3 B2 B1 B0
SYNTH 1 SYNTH 2 EXT-DC PS-DC +12 VDC -12 VDC Batt2 Batt1 Batt0
XXXXXXXXX
X = 1 is function in tolerance (ok)
X = 0 is function out of tolerance (failure)
Example:
command from the controller
response of the VOR/ILS Analyzer EVS200
query:
BI <CR> BI110111101 <CR/LF>