User manual
VOR/ILS Analyzer EVS200
0796.1800.02 E-8 14
Chapter 3: Operation
SETUP-Mode
In the following table all possible parameters are listed and its functions are described. Furthermore all
possible setting values per parameter are shown. Next a sequence chart of the operating instructions
in the SETUP mode is following.
Changes in the SETUP can be saved
with the softkey "SAVE".
parameter description setting value
RS232 BAUDRATE baud rate setting 1200, 2400, 4800,
9600, 19200
VOR ANGLE-RESOLUTION BEARING-angle resolution 0,01° / 0,05°
DDM/SDM AVG-FACTOR factor for determining the number of
measurements which form the arithmeti-
cal average
1, 2, 4, 8, 16, 32, 64
DDM Y / t-RANGE [LOC]
XY value scaling in localizer mode at
DSP-OUT (refer to page 10 DSP output)
RNG.1 to RNG.4
DDM Y / t-RANGE [GS]
XY value scaling in glideslope mode at
DSP-OUT (refer to page 10)
RNG.1 to RNG.4
AF-INPUT-SOURCE AF selection for valuation INT = internal AF
EXT = external AF
via AF-EXT-input
STARTUP mode startup mode setting at switching on the
unit
VOR (VOR mode)
ILS (ILS mode)
BCN (Beacon mode)
SPEC (Spectrum mode)
LEVEL (∆
∆∆
∆ level mode)
SQUELCH
on/off switching squelch
ON / OFF
setting valid for:
-
∆
level mode
- ILS mode
- VOR mode
When the setting is "ON" an automatical quiet tuning is performed as soon
as the threshold level < the receiver level is.
When setting is "OFF" no quiet tuning will be performed.
DISPLAY ILLUM
display brightness setting
OFF ⇐ 1to6Þ MAX
DDM/SDM DIMENSIONS value of DDM/SDM indication
µA/%/1
(1= value without dimension)
DISPLAY-INTERVALL indication interval time
setting is only valid for display indi-
cation
.
1to 8
approx. 0.4 s to approx. 1.8 s
1to8
TUNING-STEP tuning step rate setting 5 / 25 / 50 / 100 (kHz)