Datasheet

Version 05.00, December 2017
Rohde & Schwarz R&S
®
RTM Oscilloscope 19
R&S
®
RTM-K3
CAN triggering and decoding
Bus configuration
signal type
CAN_H, CAN_L
bit rate
10/20/33.3/50/83.3/100/125/250/500/
1000 kbps or user-selectable in range
from 100 bps to 5 Mbps
sampling point
10 % to 90 % within bit period
label list
associate frame identifier with symbolic ID
Trigger
trigger events
start of frame, frame type, identifier,
identifier + data, error condition (any
combination of CRC error, bit stuffing
error, form error and ACK error)
identifier setup
frame type (data, remote or both),
identifier type (11 bit or 29 bit);
condition =, ≠, >, <; identifier selectable
from label list
data setup
data pattern up to 8 byte (hex or binary);
condition =, ≠, >, <
Decode
displayed signals
bus signal, logic signal or both
color coding of bus signal
start of frame, identifier, DLC, data
payload, CRC, ACK, end of frame, error
frame, overload frame, CRC error, bit
stuffing error, ACK error
displayed format of data
hex, decimal, binary, ASCII
frame table
decode results displayed as tabulated list,
errors highlighted in red; frame navigation;
data export as CSV file
Search
search events
frame, error, identifier, identifier + data,
identifier + error
frame event setup
start of frame, end of frame, overload
frame, error frame, data ID 11 bit, data ID
29 bit, remote ID 11 bit, remote ID 29 bit
error event setup
any combination of CRC error, bit stuffing
error, form error and ACK error
identifier setup
frame type (data, remote or both),
identifier type (11 bit or 29 bit);
condition =, ≠, >, <; identifier selectable
from label list
data setup
data pattern up to 8 byte (hex or binary);
condition =, ≠, >, <
event table
search results displayed as tabulated list;
event navigation
LIN triggering and decoding
Bus configuration
version
1.3, 2.x or SAE J602; mixed traffic is
supported
bit rate
1.2/2.4/4.8/9.6/10.417/19.2 kbps or
user-selectable in range from 100 bps to
5 Mbps
polarity
active high or active low
label list
associate frame identifier with symbolic ID
Trigger
source
R&S
®
RTM3002
channel 1, channel 2, logic channels from
D15 to D0 (requires R&S
®
RTM-B1 option)
R&S
®
RTM3004
channel 1, channel 2, channel 3,
channel 4, logic channels from D15 to D0
(with R&S
®
RTM-B1 option)
trigger events
start of frame (sync break), identifier,
identifier + data, wakeup frame, error
condition (any combination of checksum
error, parity error and sync field error)
identifier setup
range from 0d to 63d; condition =, ≠, >, <;
identifier selectable from label list
data setup
data pattern up to 8 byte (hex or binary);
condition =, ≠, >, <