Datasheet
Version 05.00, December 2017
Rohde & Schwarz R&S
®
RTM Oscilloscope 5
Vertical resolution
10-bit, up to 16-bit with high resolution
decimation
DC gain accuracy
offset and position = 0
maximum operating temperature change of ±5 °C after self-alignment
input sensitivity > 5 mV/div
±1.5 %
input sensitivity ≤ 5 mV/div to
≥ 1 mV/div
±2 %
input sensitivity < 1 mV/div
±3 %
Input coupling
DC, AC, GND
Input sensitivity
at 50 Ω
0.5 mV/div to 1 V/div
at 1 MΩ
0.5 mV/div to 10 V/div
Maximum input voltage
at 50 Ω
5 V (RMS), max. 30 V (V
p
)
at 1 MΩ
300 V (RMS), 400 V (V
p
),
derates at 20 dB/decade to 5 V (RMS)
above 250 kHz
Position range
±5 div
Offset range at 50 Ω
input sensitivity
≥ 112 mV/div to 1 V/div
±(30 V – 5 div × input sensitivity)
≥ 33.8 mV/div to 111 mV/div
±(10 V – 5 div × input sensitivity)
0.5 mV/div to 33.6 mV/div
±(2 V – 5 div × input sensitivity)
Offset range at 1 MΩ
input sensitivity
≥ 515 mV/div to 10 V/div
±(250 V – 5 div × input sensitivity)
≥ 50.5 mV/div to 510 mV/div
±(25 V – 5 div × input sensitivity)
0.5 mV/div to 50 mV/div
±(2 V – 5 div × input sensitivity)
Offset accuracy
±(0.5 % × |offset| +
0.1 div × input sensitivity + 0.5 mV)
DC measurement accuracy
after adequate suppression of
measurement noise by using either high-
resolution sampling mode or waveform
averaging, or a combination of both
±(DC gain accuracy × |reading – net
offset| + offset accuracy)
Channel-to-channel isolation
(each channel at same input sensitivity)
input frequency < analog bandwidth
> 50 dB
Horizontal system
Timebase range
selectable between
0.5 ns/div and 500 s/div
Channel deskew
±500 ns
Trigger offset range
minimum
maximum
Modes
normal, roll
Channel-to-channel skew
< 200 ps (meas.)
Timebase accuracy
after delivery/calibration, at +23 °C
±2.5 ppm
during calibration interval
±3.5 ppm