Datasheet

Version 05.00, December 2017
10 Rohde & Schwarz R&S
®
RTM Oscilloscope
Mask testing
Sources
R&S
®
RTM3002
channel 1, channel 2
R&S
®
RTM3004
channel 1, channel 2, channel 3,
channel 4
Mask definition
acquired waveform with user-defined
tolerance, can be stored and restored
Result statistics
completed acquisitions, passed and failed
acquisitions (absolute and in percent),
test duration
Actions on mask violation
sound, acquisition stop, screenshot, save
waveform, pulse out (AUX OUT
connector)
Captured segments
with R&S
®
RTM-K15 option
all segments, failed segments
Waveform maths
Number of math equations
up to 5
Functions
addition, subtraction, multiplication,
division, maximum, minimum, square,
square root, absolute value, positive wave,
negative wave, reciprocal, inverse, log10,
ln, derivation, integration
Sources
R&S
®
RTM3002
channel 1, channel 2,
math waveforms 1 to 4
R&S
®
RTM3004
channel 1, channel 2, channel 3,
channel 4, math waveforms 1 to 4
Fast Fourier transformation (FFT)
Sources
R&S
®
RTM3002
channel 1, channel 2,
math waveforms, references
R&S
®
RTM3004
channel 1, channel 2, channel 3,
channel 4, math waveforms, references
Setup parameters
start frequency, stop frequency, center
frequency, frequency span, vertical scale,
vertical position, resolution bandwidth,
gate (time range and position)
Windows
Hanning, Hamming, Blackman,
rectangular, flat top
Waveform arithmetic
none, min. hold, max. hold, average
(selectable from 2 to 1024)
Search function
Functions
search types
edge, width, peak, rise/fall time, runt,
data2clock, pattern, window, protocol
(available with R&S
®
RTM-K3,
R&S
®
RTM-K6 and R&S
®
RTM-K7 options)
configuration
manual level setting on screen, level with
selectable hysteresis
display of search events
up to 10 000 events in diagram and in
result table
markers on search events
up to 32 markers
navigation in search events (stop mode)
knob (if result table is active)
Sources
R&S
®
RTM3002
channel 1, channel 2,
math waveforms from 1 to 5,
D15 to D0 (with R&S
®
RTM-B1 option)
R&S
®
RTM3004
channel 1, channel 2, channel 3,
channel 4, math waveforms from 1 to 5,
D15 to D0 (with R&S
®
RTM-B1 option)
Display characteristics
Diagram types
manually changeable vertical window size
Yt, XY, zoom, FFT, spectrogram (with
R&S
®
RTM-K18 option)
XY mode
parallel display of XY diagram and
Yt diagrams of input signals for X, Y