Datasheet
Version 05.00, December 2017
Rohde & Schwarz R&S
®
RTM Oscilloscope 9
Waveform measurements
Automatic measurements
measurements on channels,
math waveforms, reference waveforms
burst width, count positive pulses, count
negative pulses, count falling edges, count
rising edges, mean value, RMS cycle,
RMS, mean cycle, peak+, peak–,
frequency, period, amplitude, base level,
positive overshoot, negative overshoot,
pulse width, duty cycle+, duty cycle–, rise
time, fall time, delay, phase, crest factor,
slew rate+, slew rate–
reference levels
lower, middle and upper level in
percentage
statistics
maximum, minimum, mean, standard
deviation and measurement count for each
automatic measurement
number of active measurements
8
Cursor measurements
type
vertical, horizontal, vertical and horizontal,
V-marker
functions
x and y tracking, coupling of cursors, set to
trace, set to screen
Quick measurements
function
fast overview of measurements from one
channel,
some measurements displayed with result
lines in diagram
sources
R&S
®
RTM3002
channel 1, channel 2
R&S
®
RTM3004
channel 1, channel 2, channel 3,
channel 4
measurements displayed in diagram
mean, max. peak, min. peak, rise time,
fall time
numerically displayed measurements
RMS cycle, peak-to-peak voltage, period,
frequency
Digital voltmeter
Accuracy
related to channel settings of voltmeter
source
Measurements
DC, AC+DC RMS, AC RMS
Sources
R&S
®
RTM3002
channel 1, channel 2
R&S
®
RTM3004
channel 1, channel 2, channel 3,
channel 4
Number of measurements
up to 4
Resolution
up to 3 digits
Counter
Measurements
frequency, period
Sources
R&S
®
RTM3002
channel 1, channel 2, trigger signal
source
R&S
®
RTM3004
channel 1, channel 2, channel 3,
channel 4, trigger signal source
Number of measurements
2
Resolution
6 digits
Frequency range
0. 05 Hz to bandwidth of scope (limited
by bandwidth of trigger filter)