Datasheet
Version 20.00, July 2019
Rohde & Schwarz R&S
®
RTE Oscilloscope 9
Trigger system
Sources
R&S
®
RTE1022, R&S
®
RTE1032,
R&S
®
RTE1052, R&S
®
RTE1102,
R&S
®
RTE1152, R&S
®
RTE1202
channel 1, channel 2
R&S
®
RTE1024, R&S
®
RTE1034,
R&S
®
RTE1054, R&S
®
RTE1104,
R&S
®
RTE1154, R&S
®
RTE1204
channel 1, channel 2, channel 3, channel 4
Sensitivity
10
-4
div, from DC to instrument bandwidth
for all vertical scales
Trigger jitter
full-scale sine wave of frequency set to
–3 dB bandwidth
< 1 ps (RMS) (meas.)
Coupling mode
standard
same as selected channel
lowpass filter
cutoff frequency selectable from 50 kHz to
50 % of analog bandwidth
Sweep mode
auto, normal, single, n single
Event rate
max.
one event for every 800 ps time interval
Trigger level
range
±5 div from center of screen
Trigger hysteresis
modes
auto (standard) or manual
sensitivity
10
–4
div, from DC to instrument bandwidth
for all vertical scales
Holdoff range
time
100 ns to 10 s, fixed and random
events
1 event to 2 000 000 000 events
Main trigger modes
Edge
triggers on specified slope (positive, negative or either) and level
Glitch
triggers on glitches of positive, negative or either polarity that are shorter or longer than
specified width
glitch width
200 ps to 1000 s
Width
triggers on positive or negative pulse of specified width; width can be shorter, longer,
inside or outside the interval
pulse width
200 ps to 1000 s
Runt
triggers on pulse of positive, negative or either polarity that crosses one threshold but
fails to cross a second threshold before crossing the first one again; runt pulse width
can be arbitrary, shorter, longer, inside or outside the interval
runt pulse width
200 ps to 1000 s
Window
triggers when signal enters or exits a specified voltage range; triggers also when signal
stays inside or outside the voltage range for a specified period of time
Timeout
triggers when signal stays high, low or unchanged for a specified period of time
timeout
200 ps to 1000 s
Interval
triggers when time between two consecutive edges of same slope (positive or negative)
is shorter, longer, inside or outside a specified range
interval time
200 ps to 1000 s
Slew rate
triggers when the time required by a signal edge to toggle between user-defined upper
and lower voltage levels is shorter, longer, inside or outside the interval; edge slope
may be positive, negative or either
toggle time
200 ps to 1000 s
Data2clock
triggers on setup time and hold time violations between clock and data present on any
two input channels; monitored time interval may be specified by the user in the range
from –100 ns to 100 ns around a clock edge and must be at least 200 ps wide
Pattern
triggers when a logical combination (AND, NAND, OR, NOR) of the input channels
stays true for a period of time shorter, longer, inside or outside a specified range
State
triggers when a logical combination (AND, NAND, OR, NOR) of the input channels
stays true at a slope (positive, negative or either) in one selected channel
Serial pattern
triggers on serial data pattern up to 128 bit clocked by one input channel; pattern bits
may be high (H), low (L) or don’t care (X); clock edge slope may be positive, negative
or either
max. data rate
< 1.25 Gbps
TV/video
triggers on baseband analog progressive and interlaced video signals including NTSC,
PAL, PAL-M, SECAM, EDTV and HDTV broadcast standards as well as custom bi-level
and tri-level sync video standards
trigger modes
all fields, odd fields, even fields, all lines,
line number
Line
triggers with the frequency of the AC power line voltage