Datasheet

Version 20.00, July 2019
8 Rohde & Schwarz R&S
®
RTE Oscilloscope
Ultra segmented mode
continuous recording of waveforms in acquisition memory without interruption due to
visualization
max. realtime waveform acquisition
rate
> 1 600 000 waveforms/s
min. blind time between consecutive
acquisitions
< 400 ns
Differential signals
General description
Calculation of differential and common mode signals from p part and n part connected
to separate input channels. Because of the R&S
®
RTE digital trigger concept, these
signals can be used as a trigger input.
Input channels
channel 1, channel 2, channel 3,
channel 4
Differential signal
difference between two input channels
channel 1 and channel 3,
channel 2 and channel 4
Common mode signal
sum of two input channels
channel 1 and channel 3,
channel 2 and channel 4
Maximum number of outputs
differential signals
2
common mode signals
2
High definition mode
General description
The high definition mode increases the numeric resolution of the waveform signal by
using digital filtering, leading to a reduced noise. Because of the R&S
®
RTE trigger
concept, the signals with increased numeric resolution are used as input for triggering.
Numeric resolution
bandwidth
bit resolution
10 kHz to 30 MHz
16 bit
50 MHz
14 bit
100 MHz
13 bit
200 MHz
12 bit
300 MHz
11 bit
500 MHz
10 bit
Realtime sampling rate
max. 2.5 Gsample/s on each channel
RMS noise floor at 50 Ω (meas.)
input sensitivity 1 mV/div
bandwidth
10 MHz
15 µV
100 MHz
30 µV
500 MHz
61 µV
input sensitivity 10 mV/div
bandwidth
10 MHz
33 µV
100 MHz
98 µV
500 MHz
210 µV
input sensitivity 100 mV/div
bandwidth
10 MHz
250 µV
100 MHz
770 µV
500 MHz
1.65 mV