Datasheet
Version 20.00, July 2019
Rohde & Schwarz R&S
®
RTE Oscilloscope 7
Horizontal system
Timebase range
selectable between 50 ps/div and
5000 s/div,
time per div settable to any value within
range
Channel deskew
±100 ns
Reference position
0 % to 100 % of measurement display
area
Trigger offset range
max.
+(memory depth/current sampling rate)
min.
–10 000 s
Modes
normal, roll
Channel-to-channel skew
< 100 ps (meas.)
Timebase accuracy
after delivery/calibration, at +23 °C
±2 ppm
during calibration interval
±4 ppm
Delta time accuracy
corresponds to time error between two
edges on same acquisition and channel;
signal amplitude greater than 5 divisions,
measurement threshold set to 50 %,
vertical gain 10 mV/div or greater;
rise time lower than four sample periods;
waveform acquired in realtime mode
±(K/realtime sampling rate +
timebase accuracy × |reading|) (peak)
(meas.)
where
K = 0.2 (R&S
®
RTE1022, R&S
®
RTE1024)
K = 0.24 (R&S
®
RTE1032, R&S
®
RTE1034)
K = 0.27 (R&S
®
RTE1052, R&S
®
RTE1054)
K = 0.34 (R&S
®
RTE1102, R&S
®
RTE1104)
K = 0.38 (R&S
®
RTE1152, R&S
®
RTE1154)
K = 0.42 (R&S
®
RTE1202, R&S
®
RTE1204)
Acquisition system
Realtime sampling rate
max. 5 Gsample/s on each channel
Realtime waveform acquisition rate
max.
> 1 000 000 waveforms/s
Memory depth
2
R&S
®
RTE1022, R&S
®
RTE1032,
R&S
®
RTE1052, R&S
®
RTE1102,
R&S
®
RTE1152, R&S
®
RTE1202
50 Msample on 2 channels,
100 Msample on 1 channel
R&S
®
RTE1024, R&S
®
RTE1034,
R&S
®
RTE1054, R&S
®
RTE1104,
R&S
®
RTE1154, R&S
®
RTE1204
50 Msample on 4 channels,
100 Msample on 2 channels,
200 Msample on 1 channel
Decimation modes
selection valid for all channels
sample
first sample in decimation interval
peak detect
largest and smallest sample in decimation
interval
high resolution
average value of samples in decimation
interval
root mean square
root of squared average of samples in
decimation interval
Waveform arithmetic
selection valid for all channels
off
no arithmetic
envelope
envelope of acquired waveforms
average
average of acquired waveforms,
max. average depth depends on
decimation mode
3
sample
max. 16 777 215
high resolution
max. 65 535
root mean square
max. 255
reset condition
no reset (standard), manual reset
Sampling modes
realtime mode
max. sampling rate set by digitizer
interpolated time
enhancement of sampling resolution by
interpolation; max. equivalent sampling
rate is 2 Tsample/s
Interpolation modes
linear, sin(x)/x, sample&hold
2
The maximum available memory depth depends on the bit depth of the acquired data and, therefore, on the settings of the acquisition system, such as
decimation mode, waveform arithmetic, number of waveform streams or high definition mode.
3
Waveform averaging is not compatible with peak detect decimation.