Datasheet
Version 20.00, July 2019
Rohde & Schwarz R&S
®
RTE Oscilloscope 5
Base unit
Vertical system
Input channels
R&S
®
RTE1022
2 channels
R&S
®
RTE1024
4 channels
R&S
®
RTE1032
2 channels
R&S
®
RTE1034
4 channels
R&S
®
RTE1052
2 channels
R&S
®
RTE1054
4 channels
R&S
®
RTE1102
2 channels
R&S
®
RTE1104
4 channels
R&S
®
RTE1152
2 channels
R&S
®
RTE1154
4 channels
R&S
®
RTE1202
2 channels
R&S
®
RTE1204
4 channels
Input impedance
50 Ω ± 1.5 %
1 MΩ ± 1 % || 17 pF ± 1 pF (meas.)
Analog bandwidth (–3 dB)
at 50 Ω input impedance
R&S
®
RTE1022 and R&S
®
RTE1024
≥ 200 MHz
R&S
®
RTE1032 and R&S
®
RTE1034
≥ 350 MHz
R&S
®
RTE1052 and R&S
®
RTE1054
≥ 500 MHz
R&S
®
RTE1102 and R&S
®
RTE1104
≥ 1 GHz
R&S
®
RTE1152 and R&S
®
RTE1154
≥ 1.5 GHz
R&S
®
RTE1202 and R&S
®
RTE1204
≥ 2 GHz
at 1 MΩ input impedance
R&S
®
RTE1022 and R&S
®
RTE1024
≥ 200 MHz (meas.)
R&S
®
RTE1032 and R&S
®
RTE1034
≥ 350 MHz (meas.)
R&S
®
RTE1052, R&S
®
RTE1054,
R&S
®
RTE1102, R&S
®
RTE1104,
R&S
®
RTE1152, R&S
®
RTE1154,
R&S
®
RTE1202, and R&S
®
RTE1204
≥ 500 MHz (meas.)
Analog bandwidth limits
max. –1.5 dB, min. –4 dB
200 MHz, 20 MHz
Rise time/fall time
10 % to 90 % at 50 Ω (calculated)
R&S
®
RTE1022 and R&S
®
RTE1024
< 1.75 ns
R&S
®
RTE1032 and R&S
®
RTE1034
< 1 ns
R&S
®
RTE1052 and R&S
®
RTE1054
< 700 ps
R&S
®
RTE1102 and R&S
®
RTE1104
< 350 ps
R&S
®
RTE1152 and R&S
®
RTE1154
< 233 ps
R&S
®
RTE1202 and R&S
®
RTE1204
< 175 ps
Input VSWR
input frequency ≤ 500 MHz
1.25 (meas.)
input frequency > 500 MHz
1.4 (meas.)
Vertical resolution
digitizer
8 bit
high definition mode
16 bit (without reduction of the sampling
rate
1
)
high resolution decimation
16 bit (with reduction of the sampling rate)
Effective number of bits of digitizer
for full-scale sine-wave signal with
frequency equal to or lower than –3 dB
bandwidth
> 7.0 bit (meas.)
DC gain accuracy
offset and position set to 0 V, after self-alignment
input sensitivity > 5 mV/div
±1.5 %
input sensitivity ≤ 5 mV/div
±2 %
Input coupling
at 50 Ω
DC, GND
at 1 MΩ
DC, AC (> 7 Hz), GND
Input sensitivity
at 50 Ω
500 μV/div to 1 V/div,
entire analog bandwidth supported for all
input sensitivities
at 1 MΩ
500 μV/div to 10 V/div,
entire analog bandwidth supported for all
input sensitivities
1