Datasheet

Version 20.00, July 2019
36 Rohde & Schwarz R&S
®
RTE Oscilloscope
R&S
®
RTE-K50
Manchester and NRZ serial triggering and decoding
Protocol configuration
signal type
selectable,
one channel, differential or single-ended,
two channel, differential or single-ended
bit rate
auto detected, selectable/adjustable
auto threshold setup
assisted threshold configuration
source (SDATA)
analog, math. channels, logical (only NRZ)
bit encoding variants
Manchester,
Manchester II,
NRZ clocked,
NRZ unclocked
properties
active state, idle state, clock edge
frame separation
gap, enable signal (only NRZ)
Frame format
frame
multiple frame management,
frame identification and sync,
variable length frames,
variable number of cells
cells
name, size (bits), numeric format,
bit order, color
file storage of frame format
save/load as xml files
Trigger
variants
all
trigger event setup
frame start, pattern, advanced trigger
frame start
gap, start bit
pattern
up to 256 bit pattern within 65 535 bit
frame
7
advanced trigger
frame type (with OR combinations), frame
fields (with AND combinations), frame field
data; conditions =, ≠, <, ≤, >, ≥, in range,
out of range for data count, word count,
data value; error types
Decode
display type
decoded bus, logical signal, bus signal,
tabulated list, result details, decode layers
color coding
according to cell configuration table
data format
according to cell configuration table
decode layer
edges, binary
Search
search event setup
frames and frame fields, errors
Filter
The filter function selects those decode events that shall be shown in the result table.
Events that do not match the criteria set will not be displayed in the table when the filter
is turned on.
settings
same as advanced trigger settings
7
The pattern trigger will not be effective after Manchester violations.