Datasheet
Version 20.00, July 2019
30 Rohde & Schwarz R&S
®
RTE Oscilloscope
R&S
®
RTE-K7
ARINC 429 triggering and decoding
Protocol configuration
signal type
single-ended
bit rate
high (100 kbit/s)
low (12 kbit/s to 14.5 kbit/s)
polarity
A leg, B leg
device list
associate frame identifier with symbolic ID
auto threshold setup
assisted threshold configuration
timing
min. gap (0 bit to 100 bit) or off;
max. gap (0 bit to 1000 bit) or off
Trigger
trigger event setup
word start, word stop, label + data, error
condition
label + data setup
label (condition =, ≠, ≥, ≤, in range, out of
range); data (condition =, ≠, ≥, ≤, in range,
out of range); SDI/SSM
error condition
any combination of coding error, parity
error, timing error (see protocol
configuration)
Decode
source
any analog input channel, math waveform,
reference waveform
display type
decoded bus, logical signal, bus + logical
signal, tabulated list
color coding
frame (word), label, SDI, data, SSM,
parity, error condition
data format
hex, octal, binary, ASCII, signed, unsigned
Search
search event setup
word start, word stop, label + data, error
condition
event settings
same as trigger event settings
R&S
®
RTE-K8
Ethernet serial decoding
Protocol configuration
signal type
one channel, differential
bit rate
selectable/adjustable
auto threshold setup
assisted threshold configuration
source (SDATA)
analog and math channels
variants
10BASE-T, 100BASE-TX
Decode
display type
decoded bus, logical signal, bus + logical
signal, tabulated list, details, decode
layers
color coding
preamble, frame, destination address,
source address, data
data format
hex, octal, binary, signed, unsigned
decode layer
edges, binary
Search
search event setup
frame, error
frame
48 bit destination address, 48 bit source
address, 16 bit length/type, 32 bit frame
check; conditions =, ≠, <, ≤, >, ≥, in range,
out of range
error
preamble, length error