Datasheet
Version 20.00, July 2019
Rohde & Schwarz R&S
®
RTE Oscilloscope 27
R&S
®
RTE-K4
FlexRay™ serial triggering and decoding
Protocol configuration
signal type
single-ended, differential, logic
channel type
channel A, channel B
bit rate
standard bit rates (2.5/5.0/10.0 Mbps)
device list
associate frame identifier with symbolic ID
auto threshold setup
assisted threshold configuration for
FlexRay™ triggering and decoding
source
any input channel or logical channel
Trigger
trigger event setup
start of frame, header + data, symbol,
wakeup, error condition (any combination
of FSS error, BSS error, FES error, header
CRC error and frame CRC error)
header setup
indicator bits, identifier, payload length,
cycle count
indicator bits setup
payload preamble bit, null frame bit, sync
frame bit and startup frame bit separately
configurable (1, 0 or don’t care)
identifier setup
condition =, ≠, ≥, ≤, in range, out of range
payload length setup
condition =, ≠, ≥, ≤, in range, out of range
cycle count
condition =, ≠, ≥, ≤, in range, out of range;
step parameter for selection of non-
contiguous values within provided range
data setup
data pattern up to 8 byte (hex, decimal,
octal or binary); condition =, ≠, ≥, ≤, in
range, out of range; offset within frame in
range from 0 byte to 253 byte
Decode
source
any input channel, math waveform,
reference waveform, logical channel
display type
decoded bus, logical signal, bus + logical
signal, tabulated list
color coding
frame, frame header, identifier, payload
length, header CRC, cycle count, data
payload, frame CRC, error condition
data format
hex, decimal, octal, binary, ASCII
Search
search event setup
combination of start of frame, header +
data, symbol, wakeup, error condition (any
combination of FSS error, BSS error, FES
error, header CRC error and frame CRC
error)
event settings
same as trigger event settings