Version 20.00, July 2019
20 Rohde & Schwarz R&S
®
RTE Oscilloscope
R&S
®
RTE-B6
Arbitrary function/waveform generator, 2 analog channels, 8-bit pattern generator
Analog channels
General
Output channel
2 channels
Vertical resolution
14 bit
Operating modes
function generator, arbitrary waveform
generator, modulation, frequency sweep
Function generator
output of predefined waveforms
Sample rate
500 Msample/s
Waveforms
sine, square/pulse, ramp, DC, noise, sine cardinal (sinc), Gaussian pulse, Lorentz,
exponential fall, exponential rise, cardiac
Sine
frequency range
1 mHz to 100 MHz
amplitude flatness (relative to 1 kHz)
f ≤ 100 kHz
≤ ±0.1 dB
100 kHz < f ≤ 60 MHz
≤ ±0.3 dB
60 MHz < f ≤ 100 MHz
≤ ±0.5 dB
total harmonic distortion (1 V (V
pp
) into 50 Ω)
f ≤ 100 kHz
–70 dBc (=THD ≤ 0.032 %)
100 kHz < f ≤ 15 MHz
55 dBc
15 MHz < f ≤ 35 MHz
40 dBc
35 MHz < f ≤ 100 MHz
30 dBc
nonharmonic spurious (1 V (V
pp
) into 50 Ω)
65 dBc (meas.)
phase noise (meas.)
f ≤ 25 MHz
105 dBc (1 Hz) at 1 kHz offset,
115 dBc (1 Hz) at 10 kHz offset,
125 dBc (1 Hz) at 100 kHz offset
25 MHz < f ≤ 100 MHz
105 dBc (1 Hz) at 1 kHz offset,
110 dBc (1 Hz) at 10 kHz offset,
115 dBc (1 Hz) at 100 kHz offset
Square/pulse
frequency range
1 mHz to 30 MHz
duty cycle (if pulse width limit is not
exceeded)
0.01 % to 99.99 %, 0.01 % resolution
pulse width
≥ 16.5 ns, 0.1 ns resolution
rise/fall time
f ≤ 10 Hz
90 µs (meas.)
10 Hz < f ≤ 30 MHz
9 ns (meas.)
overshoot
≤ 2 %
jitter (cycle-to-cycle)
≤ 40 ps (RMS) (meas.)
Ramp (triangle, sawtooth)
frequency range
1 mHz to 1 MHz
linearity
≤ 0.1 % (meas.)
variable symmetry
0 % to 100 %, 0.1 % resolution
DC
level range
into 50 Ω
± [ 3 V (noise amplitude [V
pp
] / 2) ]
into open circuit
± [ 6 V (noise amplitude [V
pp
] / 2) ]
Noise
amplitude
DC
0 V to 6 V (V
pp
) (into 50 Ω)
0 V to 12 V (V
pp
) (into open circuit)
4 digits resolution
all other waveforms
0 % to 100 % of AC signal amplitude,
1 % resolution
bandwidth
≥ 100 MHz
Sine cardinal (sinc)
frequency range
1 mHz to 2 MHz
Gaussian pulse
frequency range
1 mHz to 10 MHz
Lorentz
frequency range
1 mHz to 5 MHz
Exponential rise/fall
frequency range
1 mHz to 1 MHz
Cardiac
frequency range
1 mHz to 1 MHz
Arbitrary waveform generator
output of user-defined waveforms
Waveform length
1 sample to 40 Msample on each channel
Sample rate
1 sample/s to 250 Msample/s
Filter bandwidth
100 MHz