Datasheet
Version 20.00, July 2019
16 Rohde & Schwarz R&S
®
RTE Oscilloscope
General data
Display
type
10.4" LC TFT color display with
touchscreen
resolution
1024 × 768 pixel (XGA)
Temperature
Temperature loading
operating temperature range
0 °C to +45 °C
storage temperature range
–40 °C to +70 °C
Climatic loading
+25° C/+40 °C at 85 % rel. humidity cyclic,
in line with IEC 60068-2-30
Altitude
Operating
up to 3000 m above sea level
Nonoperating
up to 4600 m above sea level
Mechanical resistance
Vibration
sinusoidal
5 Hz to 150 Hz, 1.8 g at 55 Hz,
0.5 g from 55 Hz to 150 Hz,
in line with EN 60068-2-6
5 Hz to 55 Hz,
in line with MIL-PRF-28800F section
4.5.5.3.2 class 3
random
10 Hz to 300 Hz,
acceleration 1.2 g (RMS),
in line with EN 60068-2-64
5 Hz to 500 Hz,
acceleration 2.058 g (RMS),
in line with MIL-PRF-28800F
section 4.5.5.3.1 class 3
Shock
40 g shock spectrum,
in line with MIL-STD-810E,
method no. 516.4, procedure I
30 g functional shock, halfsine,
duration 11 ms,
in line with MIL-PRF-28800F
section 4.5.5.4.1
EMC
RF emission
in line with CISPR 11/EN 55011 group 1
class A (for a shielded test setup);
the instrument complies with the emission
requirements stipulated by EN 55011,
EN 61326-1 and EN 61326-2-1 class A,
making the instrument suitable for use in
industrial environments
Immunity
in line with IEC/EN 61326-1 table 2,
immunity test requirements for industrial
environment
5
Certifications
VDE-GS,
C
CSA
US
Calibration interval
1 year
5
Test criterion is displayed noise level within ±1.5 div for input sensitivity of 5 mV/div.