Datasheet
Version 20.00, July 2019
Rohde & Schwarz R&S
®
RTE Oscilloscope 15
Display characteristics
Diagram types
Yt, XY, long-term measurement, spectrum, spectrogram (spectrogram requires
R&S
®
RTE-K18 option)
Display interface configuration
display area can be split up into separate diagram areas by dragging and dropping
signal icons;
each diagram area can hold any number of signals;
diagram areas may be stacked on top of each other and later accessed via the dynamic
tab menu
Signal bar
accommodates timebase settings, trigger settings and signal icons;
signal bar may be docked to left or right side of display area or hidden
Signal icon
each active waveform is represented by a separate signal icon on the signal bar; the
signal icon displays the individual vertical and acquisition settings; a waveform can be
minimized to its signal icon so that it appears as a realtime preview in miniature form;
dialog boxes and measurement results may also be minimized to a signal icon
Axis label
X-axis ticks and Y-axis ticks labeled with tick value and physical unit
Diagram label
diagrams may be individually labeled with a descriptive user-defined name
Diagram layout
grid, crosshair, axis labels and diagram label may be switched on and off separately
Persistence
50 ms to 50 s, or infinite
Zoom
user-defined zoom window provides vertical and horizontal zoom;
each diagram area supports multiple zoom windows;
touchscreen interface simplifies resize and drag operations on zoom window
Signal colors
predefined or user-defined color tables for persistence display
Input and output
Front
Channel inputs
BNC-compatible,
for details see Vertical system
probe interface
auto-detection of passive probes,
Rohde & Schwarz active probe interface
External trigger input
BNC-compatible,
for details see Trigger system
Probe compensation output
signal shape
rectangle, V
low
= 0 V, V
high
= 1 V
amplitude 1 V (V
pp
) ± 5 %
frequency
1 kHz ± 1 %
impedance
nom. 50 Ω
Ground jack
connected to ground
USB interface
2 ports, type A plug, version 2.0
Rear
Trigger out
SMA,
for details see Trigger system
USB interface
2 ports, type A plug, version 3.0
LAN interface
RJ-45 connector,
supports 10/100/1000BASE-T
External monitor interface
DVI-D connector,
output of oscilloscope display or extended
desktop display
Reference input/output
connector
BNC female,
software switch for selection of
input/output
input
impedance
nom. 50 Ω
input frequency
10 MHz
required level
≥ 0 dBm into 50 Ω
output
impedance
nom. 50 Ω
output frequency
nom. 10 MHz
level
> 7 dBm
GPIB interface
see R&S
®
RTE-B10 option
Security slot
for standard Kensington style lock