RF characteristics
4
Sensitivity/noise density
at 1.001 GHz
(measurement of the power spectral
density at 1.001 GHz at input sensitivity
1 mV/div, corresponding to 36 dBm
input range of the oscilloscope, using the
FFT with center frequency 1.001 GHz,
span 500 kHz, RBW 3 kHz)
159 dBm (1 Hz) (meas.)
at 100 kHz
(measurement of the power spectral
density at 100 kHz at input sensitivity
1 mV/div, corresponding to 36 dBm
input range of the oscilloscope, using the
FFT with center frequency 100 kHz, span
20 kHz, RBW 200 Hz)
155 dBm (1 Hz) (meas.)
Noise figure
at 1.001 GHz
(calculated based on the noise density
above)
15 dB (meas.)
at 100 kHz
(calculated based on the noise density
above)
19 dB (meas.)
Signal-to-noise ratio
measured for an input carrier with
frequency 1 GHz and level 0 dBm at input
sensitivity 70 mV/div, corresponding to
0 dBm input range of the oscilloscope,
using the FFT with center frequency
1 GHz, span 100 MHz, RBW 400 Hz at
+20 MHz from the center frequency
107 dB (meas.)
Absolute amplitude accuracy
0 to 1.5 GHz
±1 dB (meas.)
Spurious-free dynamic range
measured for an input carrier with
frequency 450 MHz and level 0 dBm at
input sensitivity 70 mV/div, corresponding
to 0 dBm input range of the oscilloscope,
using the FFT with center frequency
1 GHz, span 1 GHz, RBW 100 kHz
59 dBc (meas.)
Second-harmonic distortion
measured for an input carrier with
frequency 450 MHz and level 0 dBm at
input sensitivity 70 mV/div, corresponding
to 0 dBm input range of the oscilloscope,
using the FFT with center frequency
1 GHz, span 1 GHz, RBW 100 kHz
52 dBc (meas.)
Third-harmonic distortion
measured for an input carrier with
frequency 450 MHz and level 0 dBm at
input sensitivity 70 mV/div, corresponding
to 0 dBm input range of the oscilloscope,
using the FFT with center frequency
1 GHz, span 1 GHz, RBW 100 kHz
48 dBc (meas.)