User Manual
Remote Commands Reference
R&S
®
RTC1000
267User Manual 1335.7352.02 ─ 04
13.16.2  Parallel / Parallel Clocked Bus
BUS<b>:PARallel:DATA<m>:SOURce..............................................................................267
BUS<b>:PARallel:WIDTh................................................................................................267
BUS<b>:CPARallel:DATA<m>:SOURce........................................................................... 267
BUS<b>:CPARallel:CLOCk:SOURce............................................................................... 268
BUS<b>:CPARallel:CLOCK:SLOPe................................................................................. 268
BUS<b>:CPARallel:CS:SOURce......................................................................................268
BUS<b>:CPARallel:CS:POLarity......................................................................................269
BUS<b>:CPARallel:WIDTh..............................................................................................269
BUS<b>:PARallel:DATA<m>:SOURce <DataSource>
Defines the digital channel that is assigned to the selected bit. Use the command for
each bit of the bus.
Suffix: 
<b>
.
1, 2
Selects the bus.
<m> Selects the bit line
Parameters:
<DataSource> D0 .... D7
*RST:  D0
Example: 
BUS:PAR:WIDT 4
BUS:PAR:DATA0:SOUR D2
BUS:PAR:DATA1:SOUR D3
BUS:PAR:DATA2:SOUR D4
BUS:PAR:DATA3:SOUR D5
BUS<b>:PARallel:WIDTh <BusWidth>
Sets the number of lines to be analyzed.
Suffix: 
<b>
.
1, 2
Selects the bus.
Parameters:
<BusWidth> Range:  1 to 6
Increment:  1
*RST:  6
Default unit: Bit
BUS<b>:CPARallel:DATA<m>:SOURce <DataSource>
Defines the digital channel that is assigned to the selected bit.
Use the command for each bit of the bus.
Protocol Analysis