CURSor<m>:XDELta[:VALue]?
Returns the time difference between the two cursors (Δt).
Return values: 
<Delta> Default unit: s
Usage:  Query only
CURSor<m>:YDELta:SLOPe?
Returns the inverse value of the voltage difference - the reciprocal of the vertical dis-
tance of two horizontal cursor lines: 1/
Δ
V.
Suffix: 
<m>
.
1
The numeric suffix is irrelevant.
Return values: 
<DeltyYslope> Inverse value
Usage:  Query only
CURSor<m>:YDELta[:VALue]?
Queries the delta of the values in y-direction at the two cursors.
Suffix: 
<m>
.
1
The numeric suffix is irrelevant.
Return values: 
<DeltaY> Delta value in V
Usage:  Query only
CURsor<m>:XRATio:UNIT <Unit>
Sets the unit for X Ratio measurements with CURSor<m>:XRATio[:VALue]?.
Parameters:
<Unit> RATio | PCT | GRD | PI
RATio
floating value
PCT
percent
GRD
degree
PI
radian
*RST:  RAT