Datasheet
Version 03.00, June 2017
Rohde & Schwarz R&S
®
RTC1000 Digital Oscilloscope 5
Trigger system
Trigger level
range (min)
±15 div from center of screen
Trigger modes
auto, normal, single
Hold-off range
time
auto or 50 ns to 10 s
Trigger types
edge, pulse, video, logic, serial bus
Edge trigger
trigger events
rising edge, falling edge, both edges
sources
channel 1, channel 2, logic channels from
D7 to D0 (with R&S
®
RTC-B1 option),
external trigger input, line
coupling (analog channels, external
trigger input)
DC, AC, auto level,
low pass (attenuates > 5 kHz (meas.)),
HF (attenuates < 30 kHz (meas.)),
noise reject (enlarges trigger hysteresis)
Pulse trigger
trigger events
pulse width is smaller, greater, equal,
unequal, inside interval, outside interval
min. pulse width
8 ns
max. pulse width
17.1 s
polarity
positive, negative
sources
channel 1, channel 2, logic channels from
D7 to D0 (with R&S
®
RTC-B1 option)
Video trigger
trigger events
selectable line, all lines, even frame,
odd frame, all frames
supported standards
PAL, NTSC, SECAM, PAL-M, SDTV 576i,
HDTV 720p, HDTV 1080i, HDTV 1080p
sources
channel 1, channel 2
sync pulse polarity
positive, negative
Logic trigger
trigger events
logic condition between active channels
sources
channel 1, channel 2, logic channels from
D7 to D0 (with R&S
®
RTC-B1 option)
state of channels
high, low, don’t care
logic between channels
and/or
condition
true, false
duration condition
smaller, greater, equal, unequal, inside
interval, outside interval, timeout
min. duration time
8 ns
max. duration time
17.1 s
Serial bus trigger
supported standards
R&S
®
RTC-K1 option
I
2
C/SPI (two- and three-wire)
R&S
®
RTC-K2 option
UART/RS-232/RS-422/RS-485
R&S
®
RTC-K3 option
CAN/LIN
Trigger sensitivity
with DC, AC, LF reject
input sensitivity ≥ 5 mV/div
< 0.8 div (meas.)
input sensitivity < 5 mV/div
< 1.5 div (meas.)
with HF reject
all input sensitivities
< 1 div (meas.)
with noise rejection
input sensitivity > 5 mV/div
< 1.5 div (meas.)
External trigger input
input impedance
1 MΩ ± 1 % with 14 pF ± 2 pF (meas.)
maximum input voltage at 1 MΩ
max. 100 V (V
p
),
derates at 20 dB/decade to 5 V (RMS)
above 100 kHz
trigger level
±5 V
sensitivity
300 mV (V
pp
)
input coupling
DC, AC
Trigger output (Aux Out connector)
functionality
A pulse is generated for every acquisition
trigger event.
output voltage
at high impedance
0 V to 3.0 V
pulse polarity
high active
output delay
depends on trigger settings
pulse width
> 150 ns (trigger event)
> 0.5 µs (mask violation)