Datasheet
Version 03.00, June 2017
4 Rohde & Schwarz R&S
®
RTC1000 Digital Oscilloscope
Base unit
Vertical system
Input channels
R&S
®
RTC1002
2 channels
Input impedance
R&S
®
RTC1002
1 MΩ ± 2 % with 14 pF ± 2 pF (meas.)
Analog bandwidth (–3 dB)
R&S
®
RTC1002
> 50 MHz
R&S
®
RTC1002 with -B220 option
> 70 MHz
R&S
®
RTC1002 with -B221 option
> 100 MHz
R&S
®
RTC1002 with -B222 option
> 200 MHz (≥ 5 mV/div)
R&S
®
RTC1002 with -B223 option
> 300 MHz (≥ 5 mV/div)
Lower frequency limit (–3 dB)
at AC coupling
< 2 Hz (meas.)
Analog bandwidth limits
(max. –1.8 dB, min. –3.5 dB)
20 MHz (meas.)
Rise time (10 % to 90 %, calculated)
R&S
®
RTC1002
< 7 ns
R&S
®
RTC1002 with -B220 option
< 5 ns
R&S
®
RTC1002 with -B221 option
< 3.5 ns
R&S
®
RTC1002 with -B222 option
< 1.75 ns
R&S
®
RTC1002 with -B223 option
< 1.15 ns
Vertical resolution
8 bit, up to 16 bit with high-resolution
decimation mode
DC gain accuracy
maximum operating temperature change of ±5 °C after self-alignment
all input sensitivities
±3 % of full scale
DC measurement accuracy
after adequate suppression of
measurement noise by using high-
resolution sampling mode or waveform
averaging
±(DC gain accuracy × |reading +
sensitivity × position setting| + 0.1 div +
1 mV)
Input coupling
DC, AC, GND
Input sensitivity
1 mV/div to 10 V/div
Maximum input voltage
max. 200 V (V
p
), derates at 20 dB/decade
to 5 V (RMS) above 100 kHz
Position range
±15 div
Channel-to-channel isolation
(each channel at same input sensitivity)
input frequency < analog bandwidth
> 35 dB (meas.)
Horizontal system
Timebase range
selectable between 1 ns/div and 100 s/div
Channel deskew
±120 ns
Trigger offset range
minimum
memory depth/actual sampling rate
maximum
2
33
/actual sampling rate
Modes
normal, roll ≥ 50 ms/div
Timebase accuracy
after delivery/calibration, at +23 °C
±50 ppm
during calibration interval
±60 ppm
Acquisition system
Maximum realtime sampling rate
2 × 1 Gsample/s or 1 × 2 Gsample/s
Memory depth per channel
2 × 1 Msample or 1 × 2 Msample
Acquisition modes
refresh
first sample in decimation interval
peak detect
largest and smallest sample in decimation
interval (1 ns detection)
high resolution
average value of all samples in
decimation interval (up to 16 bit)
envelope
envelope of acquired waveforms
average
average over a series of acquired
waveforms
filter
low-pass, adjustable
smooth
Number of averaged waveforms
2 to 1024
Waveform acquisition rate
dot display, single channel,
max. waveform rate
up to 10 000 waveforms/s