Datasheet
Version 03.00, June 2017
Rohde & Schwarz R&S
®
RTC1000 Digital Oscilloscope 15
LIN triggering and decoding
Bus configuration
version
1.3, 2.x or SAE J602; mixed traffic is
supported
bit rate
1.2/2.4/4.8/9.6/10.417/19.2 kbps or user-
selectable in range from 1 kbps to 5 Mbps
polarity
active high or active low
label list
associate frame identifier with symbolic ID
Trigger
source
any input channel
trigger events
start of frame (sync break), identifier,
identifier + data, wakeup frame, error
condition (any combination of checksum
error, parity error and sync field error)
identifier setup
range from 0d to 63d; condition =, ≠, >, <;
identifier selectable from label list
data setup
data pattern up to 8 byte (hex or binary);
condition =, ≠, >, <
Decode
displayed signals
bus signal, logic signal or both
color coding of bus signal
frame, frame identifier, parity, data
payload, checksum, error condition
displayed format of data
hex, decimal, binary, ASCII
frame table
decode results displayed as tabulated list,
errors highlighted in red; three table
positions (top, bottom, full screen); frame
navigation; data export as CSV file
Search
search events
frame, error, identifier, identifier + data,
identifier + error
frame event setup
start of frame, wake up
error event setup
any combination of checksum error, parity
error and sync field error
identifier setup
range from 0d to 63d; condition =, ≠, >, <;
identifier selectable from label list
data setup
data pattern up to 8 byte (hex or binary);
condition =, ≠, >, <
event table
search results displayed as tabulated list;
event navigation