Datasheet

Version 03.00, January 2017
Rohde & Schwarz R&S
®
RTB2000 Digital Oscilloscope 11
General data
Display
Type
10.1" WXGA display with capacitive touch
Resolution
1280 × 800 pixel (WXGA)
Temperature
Temperature loading
operating temperature range
0 °C to +50 °C
storage temperature range
40 °C to +70 °C
Climatic loading
+25° C/+40 °C at 85 % rel. humidity cyclic,
in line with IEC 60068-2-30
Altitude
Operating
up to 3000 m above sea level
Nonoperating
up to 4600 m above sea level
Mechanical resistance
Vibration
sinusoidal
5 Hz to 150 Hz, max. 1.8 g at 55 Hz;
0.5 g from 55 Hz to 150 Hz,
in line with EN 60068-2-6
MIL-PRF-28800F, 4.5.5.3.2 sinusoidal
vibration, class 3 and 4
random
10 Hz to 300 Hz,
acceleration 1.2 g (RMS),
in line with EN 60068-2-64,
MIL-PRF-28800F, 4.5.5.3.1 random
vibration, class 3 and 4
Shock
40 g shock spectrum,
in line with MIL-STD-810E, method
no. 516.4, procedure I,
MIL-PRF-28800F, 4.5.5.4.1 functional
shock, 30 g, 11 ms, halfsine
EMC
RF emission
in line with EN 55011 class A, operation in
residential, commercial and business
areas or in small-size companies is not
covered; therefore the instrument may not
be operated in residential, commercial and
business areas or in small-size companies
unless additional measures are taken to
ensure that EN 55011 class B is complied
with
in line with CISPR 11/EN 55011 group 1
class A (for a shielded test setup);
the instrument complies with the emission
requirements stipulated by EN 55011,
EN 61326-1 and EN 61326-2-1 class A,
making the instrument suitable for use in
industrial environments
Immunity
in line with IEC/EN 61326-1 table 2,
immunity test requirements for industrial
environments
2
Certifications
VDE,
C
CSA
US
Calibration interval
1 year
Power supply
AC supply
100 V to 240 V at 50 Hz to 60 Hz,
0.95 A to 0.5 A
Power consumption
max. 60 W
Safety
in line with IEC 61010-1, EN 61010-1,
CAN/CSA-C22.2 No. 61010-1-04,
UL 61010-1
Mechanical data
Dimensions
W × H × D
390 mm × 220 × 152mm
(15.4 in × 8.66 in × 5.98)
Weight
(nom.)
2.5 kg (5.5 lb)
Test criterion is displayed noise level within ±1 div for input sensitivity of 5 mV/div.