Datasheet

Version 17.00, November 2016
8 Rohde & Schwarz R&S
®
RT-Zxx High-Bandwidth Probes
R&S
®
RT-ZS10L active probe
All parameters are valid when the probe is connected to an appropriate Rohde & Schwarz oscilloscope with an input impedance of 50 Ω.
See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details.
R&S
®
RT-ZS10L
Step response
Rise time
10 % to 90 %, calculated from 0.35/bandwidth
350 ps
Frequency response
Bandwidth
starting at DC
> 1 GHz (meas.)
Input impedance
DC input resistance
1 MΩ (meas.)
Input capacitance
0.9 pF (meas.)
DC characteristics
Dynamic range
derated, see figure on page 9
±8 V
Attenuation
10:1
Attenuation error
probe only, with ideal 50 Ω load impedance
±0.5 % (meas.)
Maximum rated input voltage
DC
±20 V
AC
derated, see figure on page 9
20 V (V
p
)
Base unit
Input coupling
50 Ω