Datasheet

Version 17.00, November 2016
6 Rohde & Schwarz R&S
®
RT-Zxx High-Bandwidth Probes
R&S
®
RT-ZZ80 transmission line probe
All parameters are valid for the probe only when connected to a host instrument with an input impedance of 50 Ω.
See table on page 4 and Rohde & Schwarz oscilloscope operating manual for more details.
R&S
®
RT-ZZ80
Step response
Rise time
10 % to 90 %
< 60 ps
Frequency response
Bandwidth
starting at DC
8.0 GHz (meas.)
Input impedance
DC input resistance
system
500 Ω ± 1 %
Input capacitance
0.3 pF (meas.)
DC characteristics
Attenuation
system
10:1
Attenuation error
probe only, with ideal 50 Ω load impedance
±1 %
Maximum nondestructive input voltage
Continuous voltage
20 V (RMS)
ESD tolerance
human body model
2 kV (meas.)