Remote Control
R&S
®
NGE100B
89User Manual 5601.1343.02 ─ 04
Return values:
<Power> <Numeric_value>
Return the measured power value.
Default unit: W
Example: MEAS:POW? -> 3.00E+00
Return the measured power value of the selected channel.
Usage: Query only
MEASure[:SCALar][:VOLTage][:DC]?
This command queries the measured voltage value of the previous selected channel.
Return values:
<Voltage> <numeric_value>
Return the measured voltage.
Default unit: V
Example: MEAS? -> 1.000E+00
Usage: Query only
6.5.11 Voltage and Current Configuration Commands
APPLy <arg0>
This command defines or queries the voltage and current value of the selected chan-
nel.
Parameters:
<arg0> {<Voltage> | DEF | MIN | MAX} [, {<Current> | DEF | MIN | MAX}]
<Voltage>
Set voltage in the range of 0 V to 32.20 V.
DEF
Set voltage to default.
MIN
Set voltage to minimum .
MAX
Set voltage to maximum.
<Current>
Set current in the range of 3 mA to 3.025 A.
DEF
Set current to default.
MIN
Set currernt to minimum.
MAX
Set current to maximum.
Remote Control - Commands