User Manual

Remote Control
R&S
®
NGE100B
82User Manual 5601.1343.02 ─ 04
Example:
INST OUT1
OUTP:SEL ON
Volt 12
Curr 0.1
Ch 1 is activated.
INST OUT2
OUTP:SEL ON
Volt 12
Curr 0.2
Ch 2 is activated.
OUTP:GEN ON
Ch 1 and Ch 2 and Output state are activated.
OUTP:GEN? -> 1
Return output state of the channel.
OUTPut:MODE?
Queries the output operating mode of the previous selected channel
Return values:
<arg0> {OFF | CV | CC}
OFF
Output is turned off.
CV
Output mode changed to constant voltage mode.
CC
Output mode changed to constant current mode.
*RST: OFF
Example: OUTP:MODE? -> CV
Output mode has changed to constant voltage mdoe.
Usage: Query only
OUTPut:SELect <arg0>
This command activates or queries the instrument channel state of the previous
selected channedl.
If the selected channel is activated, the respective channel key (i.e. Ch 1, Ch 2, Ch 3)
on the front panel will be illuminated.
Parameters:
<arg0> {OFF | ON | 0 | 1}
ON | 1
Selected channel will be activated.
OFF | 0
Selected channel will be deactivated.
*RST: OFF | 0
Remote Control - Commands