User Manual

Remote Control
R&S
®
NGE100B
76User Manual 5601.1343.02 ─ 04
Sets bit 0 in the event status register when all preceding commands have been execu-
ted. This bit can be used to initiate a service request. The query form writes a "1" into
the output buffer as soon as all preceding commands have been executed. This is
used for command synchronization.
*OPT?
Option identification query
Queries the options included in the instrument. For a list of all available options and
their description refer to the data sheet.
Usage: Query only
*RST
Reset
Sets the instrument to a defined default status. The default settings are indicated in the
description of commands.
Usage:
Setting only
*STB?
Status byte query
Reads the contents of the status byte in decimal form.
Usage:
Query only
*WAI
Wait to continue
Prevents servicing of the subsequent commands until all preceding commands have
been executed and all signals have settled (see also command synchronization and
*OPC).
Usage:
Event
6.5.2 System Commands
The following commands perform the various system functions and configure the com-
munication interface (LAN, WLAN) for the instrument.
SYSTem:BEEPer[:IMMediate]
The instrument returns a single beep immediately.
Usage:
Event
Remote Control - Commands