Firmware update via SCPI command for version 1.3 and above
1. Download the firmware file from the product homepage.
2. Unzip the firmware to the root directory of a USB flash drive.
There are two files after the unzip, do not rename these files.
3. Connect the USB flash drive to the USB port of the instrument and wait for ten sec-
onds.
4. Connect to the instrument via USB, LAN or WLAN. See Chapter 5.6.1.1, "Wired
Connection", on page 49 and Chapter 5.6.2, "USB Connection", on page 53.
5. Send DIAGnostic:SERVice:FWUP "NGE100FWRelease.tar.gz.enc" to
start firmware update remotely.
The R&S NGE100B reboots automatically as part of the update process.
6. Wait for the completion of the update process.
It takes a few minutes to complete the firmware udapte process.