Trigger in parameters Trigger conditions Description
Pulse Activates the EasyArb function on channel 1
when the selected pulse is met. See Chap-
ter 5.5.3, "EasyArb", on page 44.
Logic Active High Set the logic level of the trigger in signal.
Active Low
Table 5-12: Trigger out parameters and conditions
Trigger out parameters Trigger conditions Description
Channel Ch 1, Ch 2, Ch 3 Output channel selected to monitor for trig-
ger condition.
Condition Trigger Output On
Output the selected logic level when the
Output is turned on at the selected channel.
Output Off
Output the selected logic level when the
Output is turned off at the selected channel.
Fuse Tripped
Output the selected logic level when a fuse
tripped event occurs on the selected chan-
nel.
See Chapter 5.4.3, "Activating Fuse",
on page 38 and Chapter 5.5.2, "Fuse",
on page 43.
CC Mode
Output the selected logic level when the
selected channel operates in the CC mode.
See "CC mode" on page 36.
Voltage Level >= "set value" Output the selected logic level when the
voltage level of the selected channel is
greater or equal to the set level.
Current Level >= "set value" Output the selected logic level when the
current level of the selected channel is
greater or equal to the set level.
Critical Event OVP Output the selected logic level when the
selected critical event (OVP, OPP, OTP)
occurs on the selected channel. See Chap-
ter 5.5.1, "Protection", on page 41.
OPP
OTP
EasyArb Active
Output the selected logic level when the
EasyArb function is running on channel 1.
See Chapter 5.5.3, "EasyArb", on page 44.
Note: For a reliable DIO trigger output, the
EasyArb waveform needs to be at least 200
ms.
Logic Active High Set the logic level of the trigger out signal.
Active Low
1. Select "Interface > Digital IO" menu item.
The R&S NGE100B displays the "Digital IO" dialog. See Figure 5-19.
Advanced Functions