User Manual

Instrument Functions
R&S
®
NGE100B
39User Manual 5601.1343.02 ─ 04
A delay can be applied to every channel fuse. Also, individual electronic fuses (Fuse-
Link) can be logically linked so as to switch off the interlinked channels.
For more information, see Fuse for the fuse delay and FuseLink configuration.
5.4.4 Activating the Channels Output
The outputs of all the channels (Ch 1, Ch 2, Ch 3) can be switched on or off by tog-
gling the Output key on the front panel.
By default, the output is switched off when the instrument is switched on. The output is
also automatically switched off when no channel is selected. This design prevents a
connected load from being damaged unintentionally.
1. Press the desired channels.
Selected channel keys (Ch 1, Ch 2, Ch 3) illuminate.
2. Press Output key.
The R&S NGE100B output the set voltage of the selected channel.
Depending on the output mode, the font text in the channel display area shows
green when in CV mode and red when in CC mode.
See Chapter 5.3, "Output Modes", on page 35.
Figure 5-12: Output of Ch 3 in CV and CC mode
Multiple outputs can be switched on at the same time. See more information in Chap-
ter 7.1, "Parallel and Serial Mode", on page 98.
See also Power Derating.
5.4.5 Store and Recall
Upon power-up, the instrument loads the stored setting from memory location M1 with
the exception on EasyArb parameters. Auto saved parameters are also applied during
startup.
The R&S NGE100B output state of all channels (Ch 1, Ch 2, Ch 3) are disabled when
the recall function is activated.
Basic Functions