Datasheet
Version 03.00, February 2017
8 Rohde & Schwarz R&S
®
FSC Spectrum Analyzer
General data
Power supply
AC supply
input specifications
100 V AC to 240 V AC, 50 Hz to 60 Hz,
400 Hz, 130 VA
DC supply
input specifications
14 V to 16 V, 0.9 A to 0.7 A, nominal
Power consumption
12 W, nominal
Safety
in line with IEC 61010-1, EN 61010-1,
CAN/CSA C22.2 No. 61010-1-04,
UL61010-1
Test mark
VDE - GS,
C
CSA
US
,
Manual operation
Languages
Chinese, English, French, German,
Italian, Hungarian, Japanese, Korean,
Portuguese, Russian, Spanish
Remote control
Command set
SCPI 1997.0
LAN interface
10/100BaseT, RJ-45
USB interface
rear panel
USB device, type B
Display
Type
14.5 cm (5,7") LCD TFT color
Resolution
640 × 480 pixel
Audio
Speaker
internal
Mass memory
Mass memory
flash memory (internal)
USB memory stick (not supplied)
Data storage
internal
> 256 instrument settings and traces
external, on 1 Gbyte USB memory stick
> 5000 instrument settings and traces
Temperature
operating temperature range
+0 °C to +50 °C
permissible temperature range
+0 °C to +55 °C
storage temperature range
–40 °C to +70 °C
Climatic loading
relative humidity
+25/+40 °C at 85 % relative humidity
(IEC 60068-2-30)
Mechanical resistance
Vibration
sinusoidal
IEC 60068-2-6
random
IEC 60068-2-64
Shock
40 g shock spectrum,
in line with MIL-STD-810E, method 516.4
procedure 1, IEC 60068-2-27
EMC
in line with EMC Directive 2004/108/EC
including:
IEC/EN 61326-1
1,
2
,
IEC/EN 61326-2-1,
CISPR 11/EN 55011
1
IEC/EN 61000-3-2,
IEC/EN 61000-3-3
Weight and dimensions
Dimensions
W × H × D
233 mm × 158.1 mm × 350 mm
(9.2 in × 6.2 in × 13.8 in)
Weight
4.5 kg (9.9 lb)
Recommended calibration interval
1 year
1
RF emission in line with EN 55011 class A, operation in residential, commercial and business areas or in small-size companies is not covered. Thus,
the instrument may not be operated in residential, commercial and business areas or in small-size companies, unless additional measures are taken to
ensure that EN 55011 class B is complied with.
2
Immunity test requirement for industrial environment (EN 61326 table 2).